Beckman Coulter Power Express Instructions For Use Manual page 104

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.6
Bulk Inlet Error Codes (Continued)
Sensor
PS125 OFF
PS126 OFF
PS225 OFF
PS226 OFF
PS201 ON
PS202 OFF
PS202 ON
PS212 ON
PS221 ON
PS222 ON
PS223 ON
PS224 ON
PS225 ON
PS226 ON
PS231 ON
PS241 ON
PS211 ON
PS212 OFF
4-28
Problem
Passage sensor error. A
passage sensor does not
detect a carrier within a
predetermined timeout
period.
Supply box 2-1 fails to
supply.
Sample tube is detected at
sample transfer belt of left
side bin feeder.
Supply box 2-2 fails to
supply.
1.
Confirm that there is no obstacle in the PS sensor area.
2.
Confirm that the tube transfer belt is running.
3.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sample tube is in the PS201 or PS202 area
on the tube supply belt.
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sample tube is not in the area.
2.
Remove any sample tube present. If no tube is present,
clean the sensors.
3.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sample tube is in the PS211 or PS212 area
on the tube supply belt.
2.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE