Beckman Coulter Power Express Instructions For Use Manual page 90

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.3
Error Lane Error Codes (Continued)
Sensor
SL001 ON
SL001 OFF
SL002 ON
SL002 OFF
SL003 ON
SL003 OFF
SL004 ON
SL004 OFF
SL011 ON
SL011 OFF
SL012 ON
SL012 OFF
SL013 ON
SL013 OFF
4-14
Problem
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Diverter error. The diverter
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Carrier guide is not working
correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents