The Fluidics module is on the right side of the Cytometer. To access it, you must first open the
right-side cover. Refer to
Replacement/Adjustment
tubing, there is a sheath filter and a Deep Clean solution bottle. During maintenance, it may be
necessary to replace the filter (see
Adjustment
CHAPTER 11, Replacement/Adjustment
Figure 1.9 Fluidics Module View
b
PV5
PV1
1. Alarm. Emits a warning sound when there is a problem with the Fluid Container/Cubitainer capacity or
with the performance of certain operations.
NOTE
for about 30 seconds. To mute the alarm temporarily select the Mute Alerter in the status bar.
The icon disappears when the waste container/cubitainer is emptied and/or the sheath container/
cubitainer is filled/replaced.
2. Deep Clean solution peristaltic pump. Transfers cleaning solution to the flow cell.
3. Deep Clean solution bottle. Contains the diluted cleaning solution that helps to clean the flow cell.
4. Sheath fluid filter. 0.2 μm filter, for filtering sheath fluid.
1-12
Right-Side Cover Removal and Reinstallation
Procedures. Inside the module, in addition to working pumps, valves, and
Replacing the Sheath Fluid Filter
Procedures) or to add Deep Clean solution (see
CytoFLEX
PV3
PV2
c
d
e
When the alarm sounds, the Mute Alerter icon appears in the status bar. The alarm continues
Adding the Deep Clean Solution
Procedures).