10 Maintenance
10.2 Sending the device in for
Result

10.3 Changing the battery

126
EN
MEDUMAT Easy
maintenance
Risk of infection and contamination due to inadequate
hygienic safety during maintenance measures!
The device and accessories may be contaminated, and infect the
technicians or bystanders with bacteria or viruses.
 Carry out cleaning and disinfection of the device and
accessories prior to all maintenance.
 Never send in potentially contaminated devices or accessories
for maintenance work.
1. Remove the accessories
2. Clean and disinfect the device and accessories
and disinfection", page
3. Send the device and accessories to the manufacturer or
technicians who have been expressly authorized by the
manufacturer
(see "1.2 Customer Service", page
The device has been sent in for maintenance.
If you send in devices or accessories which are visibly
contaminated, they will be disposed of by the manufacturer or a
technician specifically authorized by manufacturer at your expense.
The device comes with a special 3.6 V battery.
As a rule, the battery is changed regularly within the scope of 2-
yearly maintenance and, as such, if the device is used normally, the
operator does not have to change the battery.
Nevertheless, it is recommended always to keep a spare battery
(WM 28045 LSH 14) with the device
parts", page 133)
and
CPR
(see "7 Disassembly", page
100).
(see "12.3 Replacement
(see "5 Function check", page
89).
(see "8 Cleaning
6).
51).