TYPICAL OPERATING PARAMETERS

Typical operating parameters are as follows:
Gas
Range
EXT
0 - 1000 ppm
1000 - 4500 ppm
10 - 20 %LEL
PPM
0 - 1000 ppm
1000 - 4500 ppm
LEL
0 - 10%
10 - 100%
Volume
0 - 5%
Gas
5% - 100%
Carbon
0 - 1000 ppm
Monoxide
Hydrogen 0 - 100 ppm
Sulphide
Oxygen
0 - 21%
21 - 25%
Notes:
All the values above are at normal temperature and pressure.
Humidity is between 0% and 95% RH (non-condensing).
Pressure changes at the inlet and exhaust are minimised as they
may cause transient changes in reading.
Resolution
Zero
1 ppm
5%
10 ppm
1%
5 ppm
5%
50 ppm
0.1%
+/- 0.5%
1%
0.1%
+/- 2%
1%
1 ppm
+/- 15 ppm
1 ppm
+/- 2 ppm
0.1%
+/- 0.5%
1%
N/A
Accuracy
Typically 20%
Typically 20%
2% +/- 1% LEL
1% +/- 1% Gas
+/-15 ppm + 5% reading
+/-2 ppm + 10% reading
+/- 0.5% + 3% reading
+/- 0.5% + 3% reading
A