Switching On

t
y You can now follow the brief instructions to configure further necessary
Warm-up Time
y To deliver precise results as per the Specifications, the balance must warm up
12
2
9
3
6
Turn on the device via the A key. The following appears on the display:
balance settings before beginning weighing operations.
for at least 30 minutes after initial connection to the power supply. Only then
will the device have reached the required operating temperature.
When a verified balance used in legal metrology (legal-for-trade
applications) is connected to the power, it must warm up for at least
two hours before operation.
Getting Started
®
Cubis
MSU User Manual
27