Timer-controlled Functions

Combination options:
D
®
112
Cubis
MSU User Manual
Timer-controlled Functions
Purpose:
This application is used to trigger individual balance functions automatically
at a specific time or after specific time intervals. The trigger time or interval must
occur within 24 hours (1 day).
Options:
You can select the following functions:
- Acoustic signal
- Hold display (freeze)
- Automatic printout of the display value
- Transfer of components for totalizing, formulation, or statistics
Example:
In order to determine the amount of evaporation of a sample over time, you can
place the sample on the weighing pan and print out the weight value at fixed
intervals (e. g., every two minutes).
Totalizing, statistics, formulation, mass unit conversion, second tare memory,
individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring a Task: Timer-controlled Functions
t
Define which function should be timer-controlled.
*Acoustic signal (beep tone):
Lock in readout (hold display):
on the display for the defined time period.
Automatic printout of values
at the defined time.
t
Select the mode.
*Timer set to intervals:
(e.g., every 30 sec.)
Timer set to a specific time:
(after the start of the application) (e.g., at 08:00:00).
t
Specify when the time and/or interval should be entered.
Enter fixed time:
*Time setting can be changed during process
is only entered when the task is carried out.
The balance makes a beeping sound.
The measured weight value remains
: The displayed value is printed
The function is triggered at regular intervals
The function is triggered at the fixed time
You must enter this time and/or interval now.
: The time and/or interval