Sartorius Cubis MSU116P User Manual page 5

Cubis series
Table of Contents

Advertisement

Note on cable quality:
CE conformity is only guaranteed with well shielded cables.
The shielding braid of the cable must be made of metal with at least 80% coverage and also
covered with metal foil as much as possible.
The shielding on both sides on the full metal or metalized connector casing, as well as on
the casing inlet, must be high-frequency compliant and have a flat design.
You can only turn off power to the device by pulling the plug or disconnecting the power
cord, or by unscrewing the four-pin plug at the DC supply cable for the balance.
Do not exert mechanical pressure on the display with sharp objects, as this will damage the
display. Liquid penetration can damage the electrical components.
Only use a lightly-moistened cloth for cleaning. See "Care and Maintenance" for cleaning
the device. Do not compromise the IP protection of the balance.
Do not open the balance housing. If the seal is broken, this will result in forfeiture of all
claims under the manufacturer's warranty.
The device may only be opened by specialized personnel trained by Sartorius.
Disconnect the balance from the wall outlet prior to moving the device.
Avoid condensation damage to the device. When the balance is turned on, condensation
build-up is not possible. If the device is moved from a cool environment to a warm environ-
ment, you must allow the device to acclimatize for approx. 2 hours before connecting to the
power supply.
Hazards at the Place of Installation and During Operation
3
Avoid generating static electricity and establish equipotential bonding.
A 10 kOhm ground is used when connecting the balance to the power supply. Disconnect-
ing the ground conductor is not permitted.
3
Exercise caution to avoid glass breakage.
There is a danger of lacerations or cuts caused by sharp edges on glass or metal (contact
springs), particularly when removing or replacing the glass draft shield panels; during cleaning;
if a glass sample container breaks, or when modifying the balance hardware configuration.
3
Models with a separate electronics module (e. g. MSU6.6/3.6/2.7), MSU225S/P
and MSU125P:
Do not mix up the electronics module.
Interchanging with, e. g. models of the ME/SE series will damage the device.
3
Make sure all equipment operators have received proper instructions when working with
hazardous or toxic materials.
Failure to follow correct handling procedures can result in breakage of parts or spillage of
liquids or other substances. Make sure the necessary protective clothing or equipment is
worn, such as protective gloves, clothing, eyewear.
All safety regulations applicable in the laboratory must be observed.
3
Exercise caution to avoid squeezing or crushing fingers when
– Operating the electronic draft shield
– Removing or installing the panels
– Adjusting the display
Do not expose the device to aggressive chemical vapors or to unnecessarily extreme tem-
peratures, moisture, shocks, or vibration.
The device should be set up in areas away from equipment or devices that have strong mag-
netic properties. Extreme electromagnetic fields from electrical lines should also be avoided.
Safety Instructions
®
Cubis
MSU User Manual
5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cubis msu36pCubis msu36sCubis msu66sCubis msu66p

Table of Contents