Standard Printout Parameters When the Balance is First Delivered
When the balance is first delivered, two ports are configured: The peripheral port
(Com A) is configured for one printout; the USB port (Com B) is configured for data
exchange with a PC via SBI protocol.
Printout interface: Com A (peripheral port) and Com B (USB/PC)
Protocol to Com A: Print (preset by the port configuration)
Standard printout to Com A: Standard 1 (measured value only)
GLP printing to Com A: Off, no GLP printout
Print event for output to Com A: Press the PRINT P key
Element to be output via Com A: Measured value
Protocol to Com B: SBI (preset by the port configuration)
SBI output to Com B: Single printout to the PC
Criterion for output to Com B: After stability of balance
Format for output to Com B: 22 characters
Tare after printing: Off, no tare
O
t
The cursor softkeys can be used to select configurable parameters individually.
or
t
To be guided step-by-step through the settings, use the wizard.
Example
Manual Printout to Peripheral Port, SBI Operating Mode
y After manual operation, you would like a single printout to be printed by the
connected peripheral device.
Printout interface: Com A (peripheral port)
Protocol: SBI (toggling to port configuration possible)
SBI output: Single printout
Criterion: Manual without stability
Format: 22 characters (for applications)
Tare after printing: On or off
Example Automatic Printout to Peripheral Port, SBI Operating Mode
y At a preset interval, automatic printouts should be sent to the connected
peripheral device.
Interface: Com A (peripheral port)
Protocol: SBI (toggling to port configuration possible)
SBI output: Automatic printout
Criterion: Without stability, at stability, or after weight change
Format: 22 characters (for applications)
Stop automatic printing: Not possible or by pressing Print P key.
Interval for automatic printing: Number of cycles (1, 2, 5, 10 to100) and 0.1, 0.2,
0.5 to 10 seconds
Task Management
®
Cubis
MSU User Manual
73