Statistics - Sartorius Cubis MSU116P User Manual

Cubis series
Table of Contents

Advertisement

Statistics

Purpose:
Saving and statistical analysis of weight values and other calculated values.
The values are generated as results:
– Transaction counter
– Sum of all values
– Average value
– Standard deviation
– Lowest value (minimum)
– Highest value (maximum)
– Difference between maximum and minimum
Options:
You can save a maximum of 65,535 components.
Combination options:
Density determination, calculation, averaging, weighing in percent, parts count-
ing, timer-controlled functions, checkweighing, mass unit conversion, second tare
memory, individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring a Task: Statistics
D
* = Factory setting
t
Define how the weight value will be applied.
*
Manually:
is touched.
Automatically: first value at stability:
matically as soon as the balance has stabilized after filling and the value exceeds
the minimum load.
Automatically: last value at stability
ity is applied automatically, as soon as the balance is unloaded and the value falls
short of the minimum load. If stability was reached repeatedly between measure-
ments, the last stable weight value will be stored.
You must now select the minimum load to be applied:
(
none, 10, 20, ...*100 ... 1000 digits
Weighing Mode with Automatic Storage:
t
Select the weighing mode to be used for the weight values.
Initial weighing of samples in a container.
Final readout
The weight value is applied as soon as the corresponding button
).
:
Final weighing of samples in a container.
Statistics
The weight value is applied auto-
: The last weight value with stabil-
Weighing in
®