Technical Data
Model-specific Data
Models
Readability

Weighing capacity

Tare range (subtractive)

Repeatability

Linearity
Corner load (test load [g])
Min. initial weight*
Sensitivity drift between +10 to +30°C
Typical stabilization time
Typical measurement time

External standard calibration value

(min. accuracy class)
Display result
(depending on the set filter level)
Weighing pan size (W + D)
Weighing chamber height
(draft shield DE)
Protection
Precision balances
Models
Readability
Weighing capacity
Tare range (subtractive)
Repeatability
Linearity
Corner load (test load [g])
Min. initial weight*
Sensitivity drift between +10 to +30°C
Typical stabilization time
Typical measurement time
External standard calibration value
(min. accuracy class)
Display result
(depending on the set filter level)
Weighing pan size (W + D)
Weighing chamber height
(draft shield DE)
Protection
* = Position according to OIML R76
®
188
Cubis
MSU User Manual
MSU2203S
MSU2203P
mg
1
1/10
g
2200
1010/2200
g
– 2200
– 2200
<±mg 1
1/6
<±mg 3
5
mg
2 (1000)
3 (1000)
g
1.5
1.5
±ppm/K 1
1
s
< 1
< 1
s
< 1.5
< 1.5
g
2000 (E2)
1000 (E2)
0.1 – 0.4
mm
140 + 140
mm
172

Protected against dust and water

MSU623S
MSU623P
mg
1
1/2/5
g
620
150/300/620
g
– 620
– 620
<±mg 0.7
1/2/4
<±mg 2
5
mg
2 (200)
4 (200)
g
1.5
1.5
±ppm/K 2
2
s
< 0.8
< 0.8
s
< 1
< 1
g
500 (E2)
500 (F1)
0.1 – 0.4
mm
140 + 140
mm
172
Protected against dust and water
MSU1203S
1
1200
– 1200
0.7
2
2 (500)
1.5
1.5
< 1
< 1.5
1000 (E2)
MSU323S
1
320
– 320
0.7
2
2 (200)
1.5
2
< 0.8
< 1
200 (E2)

Hide quick links: