DKD Measurement Uncertainty

Requirements:
Combination options:
D
®
116
Cubis
MSU User Manual
DKD Measurement Uncertainty
Purpose:
This application is used to ensure that the measurement uncertainty is displayed
dynamically so that it conforms to the data documented on the DKD calibration
certificate.
The function can only be used when the balance has been prepared for this before-
hand by a service technician. The service technician determines the measurement
uncertainty of the balance via an onsite DKD calibration. The technician records
the measurements and the measurement uncertainty for weighing in the DKD cali-
bration certificate. He also saves the calculated data in the balance.
Options:
The measurement uncertainty can be displayed as an absolute value (U), a relative
value (U*), or as process accuracy (PA).
Mass unit conversion, second tare memory, individual identifiers (SQmin function)
Configuring a Task: DKD Measurement Uncertainty
t
Select the measurement uncertainty display mode in terms of the maximum
capacity.
Display absolute uncertainty
Display relative uncertainty
Display process accuracy
t
If you have selected process accuracy, enter the process accuracy factor now.
: Absolute value (e. g., 12.00 g)
: Relative value (e. g., 0.000045%)
: e.g., 0.00013%