Calibration and Adjustment
®
132
Cubis
MSU User Manual
y The procedure will be executed.
t
Wait until calibration/adjustment has been completed.
y The results are displayed after the procedure is completed.
t
To return to operating mode, select
Calibration/Adjustment Using External
Calibration Weight
h
An external calibration weight is required for this function.
Please note the tolerance of the calibration weight being used.
t
If required, switch to operating mode.
t
Make sure that the weighing pan is empty.
J
t
Press
TARE
t
Select
Cal./Adj.
t
Place the calibration weight on the balance.
t
Select
External cal./adj. with factory-defined weight
option is already marked, select
y Calibration is carried out (approx. 15 to 20 sec.)
y After calibration, the determined deviation is displayed.
t
To adjust the balance, select
t
To complete calibration without adjustment, select
Back.
to tare the balance.
Start
.
.
Adjust
or, if this
Exit
.