Configure the automatic mode for the draft shield.
*
Off, no function
Close -> function -> open
while the function is executed, and then the doors re-open.
Close > function
the function is executed.
Automatic mode can be executed using the following keys:
key
TARE
key
PRINT
Cal./Adj.
softkey
or can be executed
(statistics, formulation, totalization)
h
If automatic mode is on, the draft shield doors automatically close when the
balance is switched to standby.
Assigning Opening Functions to Keys (Learning Function)
The keys should be cross-configured (for example, left key opens and closes right
door and vice versa).
The balance must be turned on.
t
Close all draft shield doors.
t
Manually operate the door you want to assign to the left key: Push the right
door open to the extent you want it to be opened (entirely or partially).
t
Press the left key
y The door will close again. Now you have configured the left key so that,
when pressed, it opens or closes the right door.
t
Proceed accordingly to configure the left door/right key.
h
Using the same method, you can assign operation of the upper door to one
of the keys. You can also configure one key to move both doors by manually
opening both doors during setup and then pressing the desired key.
Configuring the Ionizer (Optional)
If your balance is equipped with an ionizer module, you can configure the basic
settings for the ionizer module here. These settings can be changed as necessary
during any electrostatic charge eliminator task.
t
Define ionizer function.
Switch off ionizer
Manual activation with [Ion] key
on at any time by pressing the
*
Automatic execution after closing of ws
automatically, as soon as the draft shield is closed.
: Automatic mode is off.
: The draft shield doors are closed in automatic mode
: The draft shield doors are closed in automatic mode while
when a component is saved
O
.
: The ionizer is switched off.
: Ionization can be turned
Ion
key.
System Settings (Menu)
for logging applications
: Ionization is executed
®
Cubis
MSU User Manual
65