Replace the upper and side shield panels.
1. Upper draft shield panel
2. Right draft shield panel
3. Left draft shield panel
2
Removing the Display and Control Unit from Precision Balances without a Draft
Shield with a Weighing Range of up to 15 kg
t
Carefully remove all parts as shown in the illustration.
1. Weighing pan
3
2. Shield plate/Draft shield
3. Pan support
t
Keep all parts in a safe place.
t
Turn over the balance and place it on a soft surface.
t
Remove the two retaining screws.
t
Remove the display and re-insert both screws back into their holes.
t
Carefully pull the cable connected between the display and control unit from
the retainer.