System Settings (Menu)
®
46
Cubis
MSU User Manual

Leveling the Balance (Q-Level)

It is essential for exact weighing results that the balance is absolutely level.
The front leveling feet can be used to level out small tilts in the floor. Depending
on the model, the balance is equipped with either motorized or manually operated
leveling feet.
An integrated sensor detects the alignment of the balance and triggers a warning
message when leveling is required. You can configure this warning message and the
automatic leveling function (see System Settings > Balance Parameters).
When the isoCAL function is activated, the isoCal trigger is set after manual
leveling. Balances with motorized leveling feet are leveled automatically prior to
each new calibration/adjustment.
3
Note: The scale needs to be re-leveled and then adjusted each time its setup
location is changed.
When "Level me!" appears on the display, you must level the balance.
t
Make sure that the weighing pan is empty.
t
Select the option
Balance with Motorized Leveling Feet (Available for all models with a weighing
capacity of >61 g and <6200 g):
t
To start the leveling function, select
t
Follow the instructions on the display.
Balance with Manually Operated Leveling Feet
t
Follow the instructions on the display.
3
After each leveling routine, the balance must be calibrated/adjusted
(see Section "Calibration and Adjustment").
Level the balance
from the menu.
.
Next