Turn draft shield parts approx. 90 degrees to the left (loosen knurled screw).
Setting the Angle for the Display and Control Unit
The angle of the display unit can be adjusted by the user in order to ensure optimal
readability of the weight values on the display at all times.
t
The display and control unit can be tilted as desired.
Choosing a Location
t
Selecting the right setup location:
Set up the device on a stable, even surface that is not exposed to vibrations
(e.g., weighing stone).
Place the device in a location with enough free space around it so that
excessive heat cannot build up.
Maintain free access to the device at all times.
The balance may only be used in closed spaces and at a max. height above
sea level (see, chapter Techn. Data, "Ambient conditions").
Choose a location that is not subject to the following negative influences:
Heat (heater or direct sunlight)
Drafts from open windows, AC systems, and doors
Extreme vibrations during measurement
Extreme humidity
For further information, please refer to the "Proper handling of analytical balances
and microbalances" document.