Model-specific Data
Microbalances 0.001mg
Model
Readability
Weighing capacity
Tare range (subtractive)
Repeatability
Linearity
Corner load (test load [g])
Min. initial weight
Sensitivity drift between +10...+30°C
Typical stabilization time
Typical measurement time
External standard calibration value
(min. accuracy class)
Display result
(depending on the set filter level)
Weighing pan size d
Weighing chamber height
Protection
Ultra-Microbalances 0.0001mg
Model
Readability
Weighing capacity
Tare range (subtractive)
Repeatability
Linearity
1
Corner load (test load [g])
)
Min. initial weight
Sensitivity drift between zw. +10...+30°C ±ppm/K 1
Typical stabilization time
Typical measurement time
External standard calibration value
(min. accuracy class)
Display result
(depending on the set filter level)
Weighing pan size d
Weighing chamber height
Protection
1
) = Standard pan
2
) Typical min. initial weighing according to USP (United States Pharmacopeia), USP31-NF26
MSU6.6S
mg
0.001
g
6.1
g
–6.1
<±mg
0.001
<±mg
0.004
mg
4 (2 g)
mg
2
±ppm/K 1
s
< 5
s
< 8
g
5 (E2)
0.1 – 0.4
mm
30
mm
70
Protected against dust and water
MSU2.7S
mg
0,0001
g
2.1
g
–2.1
<±mg
0.00025
<±mg
0.0009
mg
2.5 (1)
mg
1
s
< 7
s
< 10
g
2 (E2)
0.1 – 0.4
mm
20
mm
70
Protected against dust and water
MSU6.6S-F
MSU3.6P
0.001
0.001/0.002/0.005
6.1
1.1/2.1/3.1
–6.1
–3.1
0.001
0.003/0.004/0.005
0.004
0.004
4 (2 g)*
5 (1 g)
4
1
1
< 5
< 5
< 8
< 8
5 (E2)
3 (E2)
0.1 – 0.4
0.1 – 0.4
50
30
70
MSU2.7S-F
0,0001
2.1
–2.1
0.00025
0.0009
2.5 (1)
1
2 (E2)
0.1 – 0.4
50
15
®