b. Activate at least four times more averages than desired during the device
measurement.
c. Press
Cal
ISOL'N STD DONE
d. Return the averaging to the original state of the measurement, and press
RESUME CAL SEQUENCE
18.To compute the error coefficients, press
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.

Enhanced Reflection Calibration

19.If you are measuring a bilateral device and want to remove the load match error,
activate the enhanced reflection calibration by pressing:
ENHANCED RESPONSE
ENH. REFL. on OFF
You can save or store the measurement correction to use for later
NOTE
measurements. Refer to
Measurement Results"
20.This completes the enhanced response correction procedure. You can connect and
measure your device under test.
RESUME CAL SEQUENCE
.
.
until ON is selected
Chapter 4 , "Printing, Plotting, and Saving
for procedures.
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
ISOLATION FWD or REV
DONE ENH RESP CAL
Cal
.
6- 25