Disassembly and Replacement
9.3.5.1

Lamp Socket

REPLACEMENT
7.
Slide the entrance slit/illumination assembly into position.
8.
Replace the two screws on top of the assembly. DO NOT OVER
TIGHTEN!
9.
Replace the monochromator cover making sure the filter wheel motor
cable is properly positioned.
10.
Replace the main circuit board (refer to "Section 9.3.1 - Main Circuit
Board").
11.
Replace the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
The lamp socket can be removed from the entrance slit/illumination
assembly after the assembly is removed from the base assembly.
However, it can also be removed without removing the entire
assembly.
REMOVAL
1.
Remove the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
2.
Disconnect the lamp cable from the main circuit board.
Figure 9.10 - Entrance Slit/Illumination Assembly
9 - 18