AGC Intensity

Min and Max AGC
Intensity Limits
Motor Speed
Thermo Fisher Scientific
The AGC Intensity screen displays the current AGC Intensity reading and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 150,000 to 300,000 Hz. If the AGC intensity reading goes
beyond either the minimum or maximum limit, an alarm is activated and
the alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
In the Main Menu, choose Alarms > AGC Intensity.
AGC INTENSITY:
ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum AGC Intensity alarm limit screen is used to change the
minimum AGC intensity alarm limit. The minimum and maximum AGC
Intensity screens function the same way.
In the Main Menu, choose Alarms > AGC Intensity > Min or Max.
AGC INTENSITY:
ACTUAL MIN: 150000 Hz
SET MIN TO: 200000 Hz
RANGE
AVG
The Motor Speed screen (read only) displays the current motor speed. A
reading other than 100.0% indicates a problem with either the motor or
the power supply.
In the Main Menu, choose Alarms > Motor Speed.
196646 Hz
150000 Hz
300000 Hz
DIAGS ALARM

INC/DEC
SAVE VALUE
DIAGS ALARM
Operation
Alarms Menu
?
Model 48i Instruction Manual
3-73