Servicing
49iQ Photometer DMC
Photometer Lamp
Replacement
7-30
49iQ Instruction Manual
Use the following procedure to remove and replace the photometer lamp.
Equipment required:
Hex drive, 3/32
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Disconnect lamp connector.
3. Unhook lamp cable from cable holder.
Figure 7–35. Lamp Removal pt 1
4. Unfasten Velcro strap.
Figure 7–36. Lamp Removal pt 2