Cleaning The Pid Lamp; Figure 7-3. Removing The Pid Sensor - Thermo Scientific TVA2020 Instruction Manual

Toxic vapor

Advertisement

Cleaning the PID
Thermo Fisher Scientific
Use the following procedure to remove and clean the PID lamp. Replace
when needed.
Lamp
WARNING Power down the instrument before performing any service
procedures.
WARNING To prevent a potential explosion, do not operate the
instrument with the PID compartment open or FID detector cap
removed.
WARNING All maintenance procedures must be performed on a clean
surface using clean tools. Avoid touching the lamp's window as well as the
metalized portion of the cell assembly with your bare fingers. Fingerprints
left on these parts may adversely affect the sensors operation. Latex gloves
are preferred, but if they are not used, your hands must be clean and free of
oils, lotions, etc. It is acceptable to hold the lamp by its glass body or by the
edges of the window.
1. Turn the instrument OFF and remove the sensor from the instrument
(Figure
Figure 7–3. Removing the PID Sensor
2. Remove the filter cap by applying slight upward pressure with the tip of
a screwdriver or X-Acto blade just below the hole in the cap and
between the cap and housing; it will pop off
7–3).
Servicing

Cleaning the PID Lamp

(Figure
7–4).
TVA2020 Instruction Manual
7-9