Min and Max Pressure

Sample Flow

Thermo Fisher Scientific
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Pressure.
PRESSURE:
ACTUAL
>MIN
MAX
RANGE
The Minimum Pressure alarm limit screen is used to change the minimum
Limits
pressure alarm limit. The minimum and maximum pressure screens
function the same way.
In the Main Menu, choose Alarms > Pressure > Min or Max.
PRESSURE:
ACTUAL MIN:
SET MIN TO:
RANGE
The Sample Flow screen displays the current sample flow reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from .350 to 1.500 L/min. If the sample flow reading goes beyond either
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Sample Flow.
SAMPLE FLOW:
ACTUAL
>MIN
MAX
RANGE
746.9 mmHg
250.0 mmHg
1000.0 mmHg
AVG
DIAGS ALARM
246.0 mmHg
500.0 mmHg ?

INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
0.750 L/min
0.350 L/min
1.500 L/min
AVG
DIAGS ALARM
Operation
Alarms Menu
Model 48i Instruction Manual
3-71