Instrument State; Instrument Action - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Instrument Action

Analog Output
Configuration
Thermo Fisher Scientific
and closes to trigger the digital input action. If nothing is connected to the
digital input pin, the state should be left at open to prevent the action from
being triggered.
Press
to toggle and set the logic state open or closed.
DIGITAL INPUT SETUP:
>LOGIC STATE

INSTRUMENT STATE

RANGE
AVG
The Instrument Action screen allows the user to choose the instrument
action that is tied to the selected digital input.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings > Select Relay > Instrument Action.
CHOOSE ACTION:
>NONE
ZERO MODE
SPAN MODE
SET BACKGROUND
CAL TO LO SPAN
AOUTS TO ZERO
AOUTS TO FS
RANGE
AVG
CAL TO HI SPAN
SCRUBBER TEST
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, voltage channels 1-6, and current
channels 1-6 (if the I/O expansion board option is installed). Configuration
choices include selecting range, setting minimum/maximum values, and
choosing signal to output.
Note The current outputs are displayed only if the I/O expansion board
option is installed.
▲ 
OPEN
DIAGS ALARM
<--
DIAGS ALARM
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Instrument Controls Menu
3-43