Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
Measurements reported in
response to DA command
Single Range Mode
Dual/Auto Range Mode
Operating and Error
D-6
Model 48i Trace Level-Enhanced Instruction Manual
Example:
Geysitech Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>0000000000 <SP>002 <SP>
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681
0000000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response if not valid in a particular
mode then a value of 0.0 will be reported.
The following measurements reported in response to DA command are for
the Model 48i Trace Level-Enhanced.
The 1 measurement reported in single range mode for the Model 48i Trace
Level-Enhanced includes:
CO
The 2 measurements reported in dual or auto range modes for the Model
48i Trace Level-Enhanced include:
low CO
high CO
See
Table D–1
Model 48i Trace Level-Enhanced.
Status
<STX>DA<CR>
Address First Concentration(E-format)=25.78
for operating status and
Address+2
Third Concentration=11.75
Table D–2
for error status for the
Thermo Fisher Scientific
Address+1