C-Link Protocol Commands

Measurement Configuration

Measurement
Configuration
B-28
Model 48i Trace Level-Enhanced Instruction Manual
This is an I series
Instrument. Screen
Information not
available
range
high range
low range
These commands report CO range in single range mode, or the high and
low ranges in dual or auto range mode. If the mode is incorrect, the
instrument responds with "can't, wrong settings". The example below
reports that the CO full-scale range is set to 50 ppm, according to
6.
Send:
range
Receive:
range 5: 5000E-2 ppm
set range Selection
set high range Selection
set low range Selection
These commands select the CO full-scale ranges, according to
The example below sets the CO full-scale range to 50 ppm.
Send:
set range 5
Receive:
set range 5 ok
Table B–6. Standard Ranges
Selection
ppm
0
1
1
2
2
5
3
10
4
20
5
50
6
100
7
200
8
500
9
1,000
10
C1
11
C2
12
C3
3
mg/m
1
2
5
10
20
50
100
200
500
1,000
C1
C2
C3
Table B–
Table
B–6.
Thermo Fisher Scientific