Measurement
Configuration
Thermo Fisher Scientific
Send:
screen
Receive:
screen
This is an I series
Instrument. Screen
Information not
available
range no
range no2
range nox
high range no
high range no2
high range nox
low range no
low range no2
low range nox
These commands report NO, NO
or the high and low ranges in dual or auto range mode. If the mode is
incorrect, the instrument responds with "can't, wrong settings". The
example below reports that the NO full-scale range is set to 50 ppb,
according to
Table
B–6.
Send:
range no
Receive:
range no 6: 5000E-2 ppm
set range Selection
set high range Selection
set low range Selection
These commands select the NO, NO
according to
Table
B–6. The example below sets the NO full-scale range to
200 ppb.
Send:
set range 5
Receive:
set range 5 ok
Table B–6. Standard Ranges
Selection
ppb
0
5
1
10
2
20
3
50
4
100
5
200
6
C1
C-Link Protocol Commands
Measurement Configuration
, and NO
range in single range mode,
2
x
, and NO
full-scale ranges,
2
x
3
μg/m
10
20
50
100
200
500
C1
Model 42i Trace Level Instruction Manual
B-31