Table 3-5. Standard Ranges - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Operation
Range Menu
3-14
Model 42i Instruction Manual
Table 3–5
lists the available operating ranges.
ranges. When switching from standard to extended ranges, the PMT
voltage must be readjusted. For more information about readjusting the
PMT voltage, see Chapter 7, "Servicing".
In the Main Menu, choose Range > NO, NO2, or NOx Range.
NO RANGE:
CURRENTLY:
SET TO:
RANGE
AVG
Table 3–5. Standard Ranges
ppb
ppm
50
0.05
100
0.10
200
0.20
500
0.50
1,000
1.00
2,000
2.00
5,000
5.00
10,000
10.00
20,000
20.00
C1
C1
C2
C2
C3
C3
Table 3–6
50
100 ?

CHANGE VALUE
SAVE VALUE
DIAGS ALARM
3
μgm
100
200
500
1,000
2,000
5,000
10,000
20,000
30,000
C1
C2
C3
lists the extended
3
mgm
0.1
0.2
0.5
1.0
2.0
5.0
10.0
20.0
30.0
C1
C2
C3
Thermo Fisher Scientific