Settings - Thermo Scientific G100 Series Operating Manual

Co2 analyser
Table of Contents

Advertisement

General Operational Instructions
Main Menu
• Rising alarms are triggered when the gas level exceeds the maximum value entered by the user.
• Falling alarms are triggered when the gas level falls below the minimum value entered by the user.
1. From the 'Main Read Screen' press the 'Menu' key on the instrument panel.
2. Press 'Key 2' to display the 'Utilities' menu.
3. Press 'Key 3' to select alarm settings and the following screen is displayed:
4. Press 'Key 1' to maintain CO

Settings

The 'Settings' option enables the operator to maintain information with regards to taking samples
and readings.
1. From the 'Main Read Screen' press the 'Menu' key on the instrument panel.
2. Press 'Key 2' to display the 'Utilities' menu.
3. Press 'Key 4' to select settings and the following screen is displayed:
4. The following instrument settings may be maintained:
1-Prompt ID:
3-Date:
4-Auto Off:
G100 Range CO
Analyser
5-6
2
Current Alarms
alarm.
2
Settings
-
Press 'Key 1' to prompt for ID code for each sample reading, answer
Yes or No accordingly.
-
Press 'Key 3' to switch the date format between dd/mm/yy and
mm/dd/yy formats.
-
Press 'Key 4' to auto switch off the instrument when not in use. Toggle
between Auto Off: Yes or No. If set to Yes, the instrument will switch
off after 10 minutes if not in use.