Table B-9. Reply Termination Formats - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Communications Configuration
B-44
Model 48i Instruction Manual
Table B–9. Reply Termination Formats
Format
00
01
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string. The following example
reports the host name is set to iSeries. This command returns "bad cmd" if
no host name has been set.
Send:
host name
Receive:
host name iSeries
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters (with no spaces). The following example sets the host name to
analyzer01. This command returns "bad cmd" if no host name has been
set.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name. The following example
reports the instrument name for the Model 48i.
Send:
instr name
Receive:
instr name
CO Analyzer
CO Analyzer
instrument id
This command reports the instrument identification (ID). The following
example reports the current setting of the instrument ID.
Send:
instrument id
Receive:
instrument id 48
set instrument id value
This command sets the instrument id to value, where value is an integer
between 0 and 127 inclusive. The following example shows the instrument
ID changed to 50.