Table B-9. Reply Termination Formats - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

C-Link Protocol Commands
Communications Configuration
B-46
Model 42i Instruction Manual
Table B–9. Reply Termination Formats
Format
00
01
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string. The following example
reports the host name is set to iSeries. This command returns "bad cmd" if
no host name has been set.
Send:
host name
Receive:
host name iSeries
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters (with no spaces). The following example sets the host name to
analyzer01. This command returns "bad cmd" if no host name has been
set.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name.
Send:
instr name
Receive:
instr name
NO-NO2-NOx Analyzer
NO-NO2-NOx Analyzer
instrument id
This command reports the instrument id.
Send:
instrument id
Receive:
instrument id 42
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Reply Termination