Download Print this page

Table B-8. Reply Termination Formats - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

C-Link Protocol Commands
Communications Configuration
B-40
Model 42i Trace Level Instruction Manual
Note When DHCP is set to on, the user-supplied addr gw, addr dns, addr
ip, and addr nm parameters are not used.
Send:
set dhcp on
Receive:
set dhcp on ok
format
This command reports the current reply termination format. The example
below shows that the reply format is 00, which means reply with no
checksum, according to
Send:
format
Receive:
format 00
set format format
This command sets the reply termination format, where format is set
according to
Table
B–8. The example below sets the reply termination
format to checksum.
Send:
set format 01
Receive:
set format 01 ok
Table B–8. Reply Termination Formats
Format
00
01
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string.
Send:
host name
Receive:
host name analyzer01
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok