C-Link Protocol Commands

Communications Configuration

Communications
Configuration
B-38
Model 42i Trace Level Instruction Manual
set time hh:mm:ss
hh = hours
mm = minutes
ss = seconds
This command sets the internal clock (24-hour time). The example below
sets the internal time to 2:15 pm.
Note If seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns address
This command sets the dns address, where address consists of four numbers
ranging from 0-255 inclusive, separated by ".".
Send:
set addr dns 192.168.1.1
Receive:
set addr dns 192.168.1.1 ok
addr gw
This command reports the default TCP/IP gateway address.
Send:
addr gw
Receive:
addr gw 192.168.1.1
set addr gw address
This command sets the default gateway address, where address consists of
four numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr gw 192.168.1.1
Receive:
set addr gw 192.168.1.1 ok
addr ip
This command reports the IP address of the analyzer.
Send:
addr ip
Receive:
addr ip 192.168.1.15
set addr ip address
This command sets the analyzer's IP address, where address consists of four
numbers ranging from 0-255 inclusive, separated by ".".
Thermo Fisher Scientific