Calibration Setup - Konica Minolta drypro 832 Service Manual

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

15.10 Calibration Setup

Set the method for performing automatic density calibration and the calibration timing.
1
Refer to
"15.1 DRYPRO 832 Utility Tool Start (Page
window of the DRYPRO 832 Utility Tool.
2
Click the [CALIBRATION SETUP] button under
The "CALIBRATION SETUP" window appears.
"CALIBRATION SETUP" window
1
2
3
4
5
The "CALIBRATION SETUP" window contains the following items:
No.
Setup Items
1
CALIBRATION
UPON FILM SET
2
SCHEDULED
CALIBRATION
3
DEFAULT CALIB.
TIME
4
ELAPSED HOURS Displays the elapsed time since last calibration.
5
[SAVE] Button
6
[BACK] Button
FILM INFO
"
Description of Settings
Set the time at which calibration is performed after loading film.
Initial Setting: NORMAL
NORMAL
Calibration is performed after registering the first queue.
AT ONCE
Calibration is performed immediately after loading film.
Set the method for automatically correcting density after a given period of time.
Initial Setting: NONE
MESSAGE
A message prompting to perform calibration after a given period of
time is displayed.
AUTO
Calibration is automatically performed after a given period of time.
NONE
Periodic calibration is not performed.
Set the time intervals at which automatic calibration is carried out.
Initial Setting: 2160
Setting range: 1-9999
Setting range: 0-9999
Save the changes and close the "CALIBRATION SETUP" window to return to the
DRYPRO 832 Utility Tool Maintenance Menu window.
Discard the changes and close the "CALIBRATION SETUP" window to return to
DRYPRO 832 Utility Tool Maintenance Menu window.
Chapter 15 DRYPRO 832 Utility Tool
15-2)", and open the maintenance menu
.
"
6
15-23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents