16.9 IMAGER MEC I/O TEST
Conduct motor and sensor operation test by operating MEC (mechanical control) when loading parts were replaced.
•
When you check each loading operation or sensor ON/OFF with the exterior cover opened,
remove the supply tray so that film that is already loaded in the supply tray won't be affected.
Caution
•
When performing work, be careful not to get your clothes or fingers caught in driving parts.
•
With the loading part driven in one position, you can drive other loading parts. However, it
should be noted that inconsistent operation could cause equipment failure.
•
After operating any loading, DRYPRO 832 may be unstable. Turn it off once and initialize it.
This section describes the procedure for displaying the "MEC I/O TEST" window.
1
Connect the service PC to the facility's network with reference to
(Page
16-3)".
2
Refer to
"16.3 How to Start the MainteTool (Page 16-9)"
The MainteTool menu window appears.
3
Click the [IMAGER MEC I/O TEST] button.
The window in which you can select device for conducting an I/O test is displayed.
4
Place a checkmark in the box of the device for conducting I/O test under "Select Target".
5
Click the [OK] button.
The "IMAGER MEC I/O TEST" window appears.
Chapter 16 MainteTool
"16.2 Setting up Network
and start the MainteTool.
16-39