Konica Minolta drypro 832 Service Manual page 370

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

Chapter 16 MainteTool
"IMAGER MEC I/O TEST" window - [Status] tab, [A/D port] tab
1
3
4
The "IMAGER MEC I/O TEST" window contains the following items:
No.
1
Sensor display
field
2
Other information
field
3
UNIT
4
LOAD
5
Test No.
6
MEC status display Displays the status of MEC.
7
ErrorNo.
8
[RUN] button
9
[STOP] button
16-40
Item
Contains tabs from [Status1] to [Status5]. Each tab displays 20 sensor names and
statuses.
For information on status name, see
16-46)".
The status of sensor, ON (detected) or OFF (undetected), is displayed .
Information on AD Port 1 to AD Port 14 is displayed in decimal number (0 - 1024).
The temperature sensor is displayed by converting to the temperature.
Select the unit for performing MEC I/O test.
For more information, see
Select load and operation.
For more information, see
Load that you can select varies, depending on the unit you choose.
Enter the test number for the I/O test.
You can enter a value from 0 to 239.
For information on test number, see
READY
MEC is ready.
BUSY
MEC is busy.
Even if you press the [RUN] button, it does not work when MEC is
busy.
If an error occurred in MEC, the error code is displayed in four digits.
Send the test number entered to DRYPRO 832 via "ID" and run a MEC I/O test
with the specified test number.
Even if you click the [RUN] button, it does not work when the MEC status display
shows "BUSY".
Stops the MEC I/O test being run.
8
9
10
Description
"
MEC I/O Test Sensor Name List (Page
"
Test Number List (Page
16-42)".
"
Test Number List (Page
16-42)".
"
Test Number List (Page
2
5
6
7
16-42)".

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents