7
Install the actuator removed in
8
Connect the connector (JP8) removed in
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
9
Install the bracket removed in
10
Install the heat process unit and left cover with reference to
9-5)"
in
"9.1.1 Removing/Installing Heat Process
11
Return the supply tray and tray shutter to their original positions.
Take sufficient care when handling the tray shutter, as it may be impossible to insert if it
becomes deformed due to rough handling.
Caution
This is the end of the process of replacing the Deodorant filter detection sensor.
Chapter 9 Replacing Loading Parts in Heat Process Unit
3
Remove the connector (JP8).
When removing the connector, do not pull by
the wires, but hold the tab on the connector
Caution
with your fingers.
Pulling by the wires too tightly can cause a
break.
4
Remove the actuator.
5
Push the tab on the back of the sensor inward
and remove the photo sensor.
6
Insert a new photo sensor in the mounting hole.
step
4.
step
3.
step
2.
Unit".
"
Installation Procedure (Page
9-17