Konica Minolta drypro 832 Service Manual page 126

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

Chapter 7 Replacing Sub Scan Unit Loading Parts
7-8
3
Install the ground wire.
• One screw (M4×8)×2
4
Connect the connectors (CCN6) and (JJ25).
Be sure to push the connector all the way in
when connecting it.
Caution
If the connector is not connected properly,
an error will occur and, in some cases,
smoke and/or an odor may be generated.
5
Remove the connectors (JJ15) and (JJ18).
Be sure to push the connector all the way in
when connecting it.
Caution
If the connector is not connected properly,
an error will occur and, in some cases,
smoke and/or an odor may be generated.
6
Install the reference guide.
• Four screws (M3×6)
When installing the reference guide, be
careful not to come into contact with the
Caution
position regulation sensor.
7
Install the guide unit assembly by hooking the
stop bands in two locations on the external guide
unit assembly.