Chapter 15 DRYPRO 832 Utility Tool
No.
Setup Items
1
NORMAL FILM
2
3
4
HIGH DENSITY
FILM
5
6
4
Click the [BACK] button.
The "DENSITOMETER ADJUST" window closes, returning to the Maintenance Menu window.
This completes the process of checking the set density value.
15-50
Description of
Settings
INTERNAL
The results of measuring with a built-in densitometer are
shown, when a correction is made to BLUE film or CLEAR
film.
EXTERNAL
The results of measuring with an external densitometer are
shown, when a correction is made to BLUE film or CLEAR
film.
RESULT
If the result of comparing the internal density value with the
external density value is within ± 3, "OK" will be shown; or
if it is ± 4 or greater, "NG" will be shown.
INTERNAL
The results of measuring with a built-in densitometer are
shown, when a correction is made to HIGH BLUE film.
EXTERNAL
The results of measuring with an external densitometer are
shown, when a correction is made to HIGH BLUE film.
RESULT
If the result of comparing the internal density value with the
external density value is within ± 3, "OK" will be shown; or
if it is ± 4 or greater, "NG" will be shown.
Description of Settings