Chapter 7 Replacing Sub Scan Unit Loading Parts
4
Open the front cover.
The front cover will open by pulling its upper part on both
sides.
Do not push down or place objects on the
opened front cover. Failure to observe this
Caution
caution can cause the breakdown.
5
Open the external guide assy by rotating the lever
B in the direction of an arrow.
6
Open the external guide unit assembly after
removing the stop band at two places.
7
Remove the guide unit assembly.
• Four screws (M3×6)
If the surface of the guide is scratched, it
could scratch film during transportation of
Caution
film. When removing or installing the guide,
be careful not to scratch the surface of the
reference guide.