Konica Minolta drypro 832 Service Manual page 92

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

Chapter 5 Replacing Tray Unit Loading Parts
15
Connect the connectors removed in
If the tray unit tray detection sensor has been replaced, connect the connector (JP55, JP44).
If the Lispl-832 tray detection sensor has been replaced, connect the connector (JP63, JP44).
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
16
Install the connector cover removed in
• One screw (M3×6)
If connecting the Lispl-832, connect connectors (JP16, JP 26).
Note
If the tray unit tray detection sensor has been replaced, proceed to
If the Lispl-832 tray detection sensor has been replaced, proceed to
Note
5-18
11
Remove the tray lock solenoid unit assembly.
• Three screws (M3×6)
The bracket is hooked on the notch in the main unit, but can
be unhooked by sliding slightly to the right and pulling back.
12
Push the tab on the back of the sensor inward
and remove the photo sensor.
13
Insert a new photo sensor in the mounting hole.
14
Install the tray lock solenoid unit assembly
removed in
step
11.
• Three screws (M3×6)
After inserting the bracket notch into the notch groove of the
main unit (
in figure on left), align the position of the other
notches and attach.
step
10.
step 7
or
step
9.
step
17.
step
18.