Chapter 7 Replacing Sub Scan Unit Loading Parts
13
Install the gears removed in
14
Install the set screw removed in
• One set screw (M3×6)
15
Install the bracket removed in
• Two screws (M3×6)
16
Connect the connectors (JP58) and (JP35) removed in
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
17
Install the steel plate removed in
18
Connect the connector (JJ51) removed in
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
19
Install the exposure unit mounting frame unit assembly removed in
• Four screws (M4×6)
20
Connect the connector (JP10) removed in
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
7-32
8
Remove the bracket.
• Two screws (M3×6)
9
Remove the set screw.
• One set screw (M3×6)
10
Remove the gears.
11
Remove the Sub Scan nip motor.
• Two screws (M3×4)
12
Install a new Sub Scan nip motor.
• Two screws (M3×4)