18.4 Film Management Information

18.4.1 Management Information Print Specifications

Stamp writing contents
No.
Item
1
LUT reception
2
Contrast
3
Density
4
Interpolation
No
5
Polarity
6
Maximum
density
7
Minimum
density
8
Total number
of films
9
Version
10
SERIAL NO.
11
CH Number
12
SCU name
13
PrintIndexNo
14
Page number
15
DICOM
Directory
16
Requested
image size
operation
17
Requested
image size
Max.
Format
number of
characters
"G"
1 character
%1d
% 2d
2 characters
Received (2011,1031) contrast value (-7~7)
% 2d
2 characters
Received (2011,1030) density value (-7~7)
%1d
1 character
Received (2010,0080) Correction number (1~7) requested
from flat smooth type.
"N"
1 character
"P"
%3d
3 characters
Received (2010,0130) maximum density
%3d
3 characters
Received (2010,0120) minimum density
%6d
6 characters
Number of films output from the server PC (count-up from
print end command)
%-11.11s
11 characters Version of server PC installed in Drypro 832
%-5.5s
5 characters
Installed Drypro 832 main unit serial number
%-2.2s
2 characters
xx part of received (2011.1060) modality ID
"xx:yyyyyyyyyyyy"
%-12.12s
12 characters yyyyyyyyyyyy part of received (2011.1060) modality ID
"xx:yyyyyyyyyyyy"
%03d
3 characters
Print management information
%03d
3 characters
Page number
%-15.15s
15 characters Name of directory that stores DICOM received data that
DCM created
" "
1 character
"D"
"R"
"000.00"
6 characters
"000.00" when unspecified.
"Xxx.xx"
Prints the specified size (e.g. "200.00")
Chapter 18 Technical Information
Remarks
"G" when the Presentation LUT data reception flag is
received.
The received LUT number (1~7) when the Presentation
LUT data reception flag is not received.
"N" when received (2020,0020) polarity is
"NORMAL" and received (0028,0004) optical
measurement construction is "MONOCHROME1"
"P" when received (2020,0020) polarity is
"NORMAL" and received (0028,0004) optical
measurement construction is "MONOCHROME2"
"P" when received (2020,0020) polarity is
"REVERSE", and received (0028,0004) optical
measurement construction is "MONOCHROME1"
"N" when received (2020,0020) polarity is
"REVERSE" and received (0028,0004) optical
measurement construction is "MONOCHROME2".
(2020,0030) " " when requested image size is not
received
(2020,0030) "D" when the requested image size is
received but the requested image size cannot be output
(2020,0030) "R" when the requested image size is
received, and the requested image size can be output
18-21