15.20.2 Confirming After Density Calibration

Check the set density value by comparing the reading given by the densitometer built into DRYPRO 832 with the
measured value of the densitometer calibration pattern entered in the Densitometer Calibration window.
1
Click the [DENSITOMETER] button under "SERVICE INFO."
The "DENSITOMETER" window appears.
2
Click the [DENSITOMETER ADJUST] button.
The "DENSITOMETER ADJUST" window appears.
3
Check the results by comparing the internal density values with the external density values
from the (01)st to the (12)th.
If a correction was made to BLUE film or CLEAR film, check the results field on the "NORMAL FILM" side; or if
a correction was made to HIGH BLUE film, check the results field on the "HIGH DENSITY FILM" side.
(If the difference in the (01)st to the (12)th values is within ±3, the result of correction is judged as being
appropriate, and "OK" will be shown in the results column.
If the value difference is ±4 or greater, NG will be shown in the results column. Perform
"15.20.1 Built-in
Densitometer Calibration (Page 15-42)"
again.
15-49