Chapter 17 Troubleshooting

17.2 Error display and solutions

When an error occurs, an error message and the solution are displayed in the message display window of the
operating panel.
The error message is also displayed inPrintlink5-IN.

17.2.1 Operating panel display and solutions

Error
Error
Error
Message (Display
Level
Code
Characters)
E
3000
HPRO Error
F
3001
HPRO Error
E
3002
Temperature error
E
3003
Temperature error
E
3004
Temperature error
F
3006
HPRO Error
17-14
Error contents
Error when 22 minutes, which is 2times
stronger than normal startup 11 minutes
(when not in POWER OFF MODE),
passes
Check the sensor detection of wire
malfunction signal for the heater every
0.1 seconds, and an error occurs if
anomaly is detected
When the pre-heating C1 30 second
moving average temperature is compared
to 100°C, an error occurs when the
temperature is 5°C higher when Ready
and 10°C higher when in PRINT.
When the pre-heating C1 30 second
moving average temperature is compared
to 100°C, an error occurs when the
temperature is 5°C lower when Ready
and 10°C lower when in PRINT.
Irrespective of the status of the machine,
check the temperature of preheat C1
every 1 seconds, and if the temperature is
continually 147°C for 5 seconds, an error
occurs
The heater sensor A/D value has a fixed
upper value
Error Support: Service
1.
Check whether the short-circuit breaker (domestic) or circuit
protector (domestic) is OFF
2.
Check the connection between the H-DRVboard and mechanical
control board
3.
Check conductance of the power cable
4.
Check voltage
5.
Replace H-DRV board
6.
Replace heat process unit
7.
Replace mechanical control board
8.
Replace the wire bundle between the H-DRVboard and
mechanical control board
1.
Check connection between mechanical control and HPRO unit
2.
Replace HPRO unit
3.
Replace mechanical control board
4.
Replace wire bundle between mechanical control board and HPRO unit
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and mechanical
control board
3.
Check connection between mechanical control board and HPRO
unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard and
mechanical control board
10. Replace wire bundle between mechanical control board and HPRO unit
1.
Check HPRO OFFSET and mechanical control offset
2.
Check whether the short-circuit breaker (domestic) or circuit
protector (domestic) is OFF
3.
Check the connection between the H-DRVboard and mechanical
control board
4.
Check connection between mechanical control board and HPRO
unit
5.
Check conductance of the power cable
6.
Check voltage
7.
Replace H-DRV board
8.
Replace heat process unit
9.
Replace mechanical control board
10. Replace the wire bundle between the H-DRVboard and
mechanical control board
11. Replace wire bundle between mechanical control board and HPRO unit
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and mechanical
control board
3.
Check connection between mechanical control board and HPRO
unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard and
mechanical control board
10. Replace wire bundle between mechanical control board and
HPRO unit
1.
Replace HPRO unit
2.
Replace mechanical control board
3.
Replace wire bundle between mechanical control board and HPRO unit