Chapter 7 Replacing Sub Scan Unit Loading Parts
7-6
8
Remove the fixing screws of sub scan unit.
• Four screws (M4×8)
9
Remove the connectors (JJ15) and (JJ18).
When removing the connector, do not pull by
the wires, but hold the tab on the connector
Caution
with your fingers.
Pulling by the wires too tightly can cause a
break.
Turn the wire disconnected from the connector
toward the main unit to keep out of the way of
Note
the sub scan unit when it is pulled out.
10
Remove wire from the locking wire saddle and
take off the connectors (CCN6) and (JJ25).
When removing the connector, do not pull by
the wires, but hold the tab on the connector
Caution
with your fingers.
Pulling by the wires too tightly can cause a
break.
Turn the wire disconnected from the connector
toward the main unit to keep out of the way of
Note
the sub scan unit when it is pulled out.
11
Remove the ground wire.
• One screw (M4×8)×2