The formula of resolution of Z is as follows;
FWHM : Resolution of Z
L : The length of a diagonal line of Square Pinhole
NA = nSINθ :NA of Objective lens
MG : The objective Magnification
N : The refractive index of a medium
Air : 1.0
Water : 1.3
Oil : 1.5
X = L/(1.22*3.8*λ*MG/NA)
The case where the variable x is as follows; 0 <= x < 2.0
FWHM = (2λ*ZPSF)/(4n(1-COSθ))
-2
5
ZPSF = -5.5158*10
x
+ 2.0877*10
-1
6.9038*10
The case where the variable x is as follows; 2.0 <= x
FWHM = (2λ*ZPSF)/(4n(1-COSθ))
ZPSF = 0.6164x+0.1283
Getting Started FLUOVIEW
One Point!
-1
4
-1
3
x
- 2.1172*10
x
+ 2.0864*10
Page
/Online Help
-1
2
-2
x
- 2.2669*10
x +
OPERATION INSTRUCTIONS