Olympus FLUOVIEW FV1000 User Manual page 245

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Command
Entry form
Filter [NORMAL
Filter
/KALMAN/PEAK]
[ ,value]
Gain ch, value
Gain
LivePlot
LivePlot
[ON/OFF]
Merge
Merge
MergeFile1,
MergeFile2
[ ,Ch(s)][ ,Ch(s)]
Message message
Message
APPLIED OPERATIONS
Description
designated is to be Valid, select ON and, in
case that Invalid is to be set, select OFF.
<E.g> In case that channel 1 is to be Valid,
enter [EnCh1, ON].
[NORMAL/KALMAN/PEAK]: It defines type of
integration.
NORMAL: No integration
KALMAN: KALMAN filter
PEAK: Peak detecting integration
value: In case that KALMAN is selected,
integration cycle of KALMAN filter is set and,
in case of PEAK, max. addition cycle at PEAK
integration is set. In case that NORMAL is
selected, no setting is required.
<E.g> In case that integration is required 4
times, using KALMAN filter, enter [Filter
KALMAN, 4].
Sets the Gain value of channel.
<E.g.> Enter "Gain 2, 50" to set 50 to Gain of
channel 2.
[LivePlot
ON]
is
selected,
displayed.
In case of
[LivePlot OFF], LivePlot is not
displayed. (This is TIEMPO option.)
Merges image data that is located under
FLUOVIEW software.
The following example is to merge 2ch of File1
and 3ch of File2.
<E.g>
[Merge C:¥FLUOVIEW¥Images
¥File1,C:¥FLUOVIEW¥Images¥File2,2,3]
Displays a message to suspend executing a
OPERATION INSTRUCTIONS
Page
/Protocol processor
LivePlot
is
2 - 1 6 1

Advertisement

Table of Contents
loading

Table of Contents