Advertisement

8.
Set the parfocality correction values of objectives by repeating steps 6 and 7 for each.
The [Pfcl Setup Wizard] group box displays the option button menu shown below.
Click the <Pfcl> button to activate the parfocality correction.
Click to save the currently set
parfocality
correction
and
sensitivity values, then close the
dialog box.
(This option button is displayed when
the setting is changed.)
(Click the <Next> button to save the
setting and close the dialog box.)
9.
Set the jog sensitivity values of the objectives.
Select the jog sensitivity of each objective from the [Jog Fine] drop-down list.
The drop-down list appears when clicking each value in the [Jog Fine].
10. After completing the setting, click the [Save each Perfocal value and Jog Setting, then
Exit] option button and then click the <Next> button.
The setting will be registered and the dialog box will close.
If you do not want to register the setting, simply click the <Exit> button to close the
dialog box.
Getting Started FLUOVIEW
jog
/Online Help
[Jog Fine] group box
Select the jog sensitivity
of each objective from
the drop-down list.
The * mark represents
the value recommended.
OPERATION INSTRUCTIONS